XENEX CO., LTD.
ECG Analysis System

ECG Analysis System

Send Inquiry

Specification:

Display

LCD Type : 5.7", Mono
Resolution : 320 x 240 dots

Computerized
Electrocardiograph

ECG Storage - Internal : 50, External : 5,000
Acquisition & Analysis : 12 LEADS Simultaneously
Dynamic Range : AC differential : ±5mV, DC offset : ±300mV
Sampling Rage(Samples/Second/Channel) : 1,000
Frequency Response(Hz) : 0.05 to 150
Common Mode Rejection(dB) : over 100
Input Impedance(MΩ) : Over 50
Communication : RS232 or RJ45
Firmware Upload : Use the USB Memory Stick
Filte(Hz)
AC line: 50, 60
Low Pass : 40, 100, 150
High Pass : 0.05, 0.16, 0.3, 0.5
Muscle : O
Anti-Floating System(AFS) : O

Writer

Type : Thermal dot arrary(Voltage axis : 8dots/mm, time axis : 16dots/mm)
Number of Traces : 3.6 or 12 user-selectable
Sensitivity / Gain : 5, 10, 20mm/mV
Speeds : 5, 10, 25, 50mm/sec
Amplitude Accuracy : ±5%
Speed Accuracy : ±5%
Paper - Z-fold(W x L x sheets) : 282mm x 215mm x 200 sheets

Electrical

Power Supply : AC or Battery operation
AC Input : 100~240VAC 50/60Hz, 110VA
Battery Type : 14.4V Rechargeable Ni-MH, 3,800mAh
Battery Capacity - Printing at Automode : 200 records
Standby Time : 5 hours

Physical

Dimension(W x L x H) : 385mm x 367mm x 101mm
Weight : 4.5kg

Safety

Classification : Class 1, Internally powered
Type of protection : Type CF with defibrillation protection
Patient Leakage Current : below 10µA